| Model Number |
CD3231-40 |
| Medical Device Problem Codes |
Break (1069); Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported, a patient's implantable cardioverter defibrillator (icd) presented in backup vvi mode.For an unknown reason.The internal components were also found damaged.And a reset could not be performed.The device was explanted in a procedure on (b)(6) 2023.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of backup operation was confirmed.Analysis of device image found the device went into backup operation due to a power-on-reset (por).After the device was restored from backup mode, functional testing was performed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.The backup operation could not be reproduced.The cause of the reported event could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h6.
|
| |
|