| Model Number |
8884721252E |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the doctor was inserting the nasogastric tube and when he went to remove the stylet the purple end came off and the wire struck the doctor in the finger piercing the glove.The doctor was ok, no injury occurred.Additional information received on 21aug2023 stated that the nasogastric tube was not flushed with water prior to stylet removal.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information available to us, we were unable to confirm the event as the device history record was reviewed and indicated that the product was released accomplishing all quality standards.Furthermore, the reported issue could not be confirmed from the returned sample.No corrective actions are required at this time.In the meantime, all information received will be used for further tracking and trending purposes.
|
| |
|