CORDIS US CORP. SABER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number 48005010S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
As reported, a 5mm x 10cm saber percutaneous transluminal angioplasty (pta) dilation catheter ruptured at eight (8) atmospheres (atm).An alternative saber was used, and the procedure was completed successfully.There was no reported patient injury.A non-cordis 0.014 guidewire was used for support in the vascular access intervention therapy (vaivt) and passed.Additional event details were requested; however, could not be obtained.The device will not be returned for evaluation.
 
Additional Manufacturer Narrative
As reported, a 5mm x 10cm saber percutaneous transluminal angioplasty (pta) dilation catheter ruptured at eight atmospheres (atm).An alternative saber was used, and the procedure was completed successfully.There was no reported patient injury.A non-cordis 0.014 guidewire was used for support in the vascular access intervention therapy (vaivt) and passed.Additional event details were requested; however, could not be obtained.The product was not returned for analysis.A product history record (phr) review of lot 82275993 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon- burst at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown, and procedural factors likely contributed to the reported event, as damage to balloon material may have occurred during crossing or upon inflation.However, without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS US CORP.
14201 nw 60th avenue
miami lakes, florida 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes, florida 33014
Manufacturer Contact
karla castro
coyol free zone, building b25
el coyol alajuela bld 25, costa rica CR 20-102
CS   CR 20102
7863138372
MDR Report Key17633441
Report Number3007635982-2023-00273
Device Sequence Number12685653
Product Code LIT
UDI-Device Identifier20705032068625
UDI-Public(01)20705032068625(17)260228(10)82275993
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue Number48005010S
Device Lot Number82275993
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2023
Initial Report FDA Received Date08/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SABERX5MM 10CM; UNK BASSARO SUPPORT 0.014
Patient SexUnknown
-
-