U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY C CREATININE REAGENT KIT; ALKALINE PICRATE, COLORIMETRY, CREATININE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY C CREATININE REAGENT KIT; ALKALINE PICRATE, COLORIMETRY, CREATININE Back to Search Results
Catalog Number 07P99-20
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer stated that falsely elevated alinity c creatinine results were generated.The following data was provided.Sid (b)(6): initial result of 347.0 umol/l retested at 73.2 umol/l after recentrifugation.Sid (b)(6): initial result of 212.2 umol/l retested at 55.7 umol/l after recentrifugation.No impact to patient management was reported.
 
Additional Manufacturer Narrative
After further evaluation, the suspect medical device was updated on september 6, 2023.Mdr number 3016438761-2023-00515 has been submitted for the new suspect medical device and all further information will be documented under that mdr number refer to section h10.
 
Event or Problem Description
The customer stated that falsely elevated alinity c creatinine results were generated.The following data was provided.Sid (b)(6): initial result of 347.0 umol/l retested at 73.2 umol/l after recentrifugation sid (b)(6): initial result of 212.2 umol/l retested at 55.7 umol/l after recentrifugation no impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C CREATININE REAGENT KIT
Common Device Name
ALKALINE PICRATE, COLORIMETRY, CREATININE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key17633802
Report Number3002809144-2023-00369
Device Sequence Number6473698
Product Code CGX
UDI-Device Identifier00380740135102
UDI-Public00380740135102
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K083809
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2024
Device Catalogue Number07P99-20
Device Lot Number53990UN22
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2023
Supplement Date Received by Manufacturer09/06/2023
Initial Report FDA Received Date08/28/2023
Supplement Report FDA Received Date09/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/30/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, (B)(6).; ALNTY C PROCESSING MODU, 03R67-01, (B)(6).
Patient SexUnknown
-
-