| Catalog Number |
07P99-20 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer stated that falsely elevated alinity c creatinine results were generated.The following data was provided.Sid (b)(6): initial result of 347.0 umol/l retested at 73.2 umol/l after recentrifugation.Sid (b)(6): initial result of 212.2 umol/l retested at 55.7 umol/l after recentrifugation.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation, the suspect medical device was updated on september 6, 2023.Mdr number 3016438761-2023-00515 has been submitted for the new suspect medical device and all further information will be documented under that mdr number refer to section h10.
|
| |
|
Event or Problem Description
|
|
The customer stated that falsely elevated alinity c creatinine results were generated.The following data was provided.Sid (b)(6): initial result of 347.0 umol/l retested at 73.2 umol/l after recentrifugation sid (b)(6): initial result of 212.2 umol/l retested at 55.7 umol/l after recentrifugation no impact to patient management was reported.
|
| |
|