DEPUY ORTHOPAEDICS INC US ENDUR DIAMOND TOOTH BROACH SZ3; HIP INSTRUMENTS : BROACHES Back to Search Results
Catalog Number 252103510
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the sz 1 replacement broaches arrived and were a different design.Neck segments were too loose on the new broaches.Unknown surgical delay.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary: it was reported that the sz 1 replacement broach arrived today.It is a totally different design, so all of the broaches need to be changed out.Replacement endurance numbers are given.The cdh is not listed and is likely obsolete.Neck segments are too loose on the new broaches, so they need to be replaced as well.Unknown surgical delay.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the endur diamond tooth broach sz3 post is deformed and also present signs of scratches.Additionally, the device exhibits signs of heavy use causing the etching to be unreadable.The potential cause for the deformation at the post can be attributed to unintended use error, with the damage likely resulting from off-axis assembly, prying motion while device is assembled and impaction forces combined with handle not being fully secured onto the broach.Proper handling and adherence to the approved use of the device can diminish the risk of failure.The observed illegible etch condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was unable to be performed due to the post-manufacturing damage.A dimensional inspection was not performed since it was not applicable to the complaint condition.A manufacturing records evaluation (mre) was not performed since the lot number was not able to be retrieved for the device due to the illegible etch.The overall complaint was confirmed as the observed condition of the endur diamond tooth broach sz3 would have contributed to a device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing records evaluation (mre) was not performed since the lot number was not able to be retrieved for the device due to the illegible etch.Added: d9.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint #
=
> pc-001422484 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: corrected: d4 (primary udi number) if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDUR DIAMOND TOOTH BROACH SZ3
Common Device Name
HIP INSTRUMENTS : BROACHES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17634501
Report Number1818910-2023-17673
Device Sequence Number7691776
Product Code HTQ
UDI-Device Identifier10603295127925
UDI-Public(01)10603295127925
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number252103510
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2023
Supplement Date Received by Manufacturer09/19/2023
04/04/2025
04/29/2025
Initial Report FDA Received Date08/28/2023
Supplement Report FDA Received Date09/19/2023
04/09/2025
04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-