• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 32337
Medical Device Problem Codes Break (1069); Defective Device (2588)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/03/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was pulled and forced out with extreme pain to patient.An 8fr x 35cm flexima drainage catheter was used for inserting internal and external biliary drain.During the procedure, after placing the drain, the plastic inner stylet would not come out and eventually broke.The physician decided to cut the catheter hoping that would release the pigtail on the end of the catheter, but it did not help.The device was pulled and forced out which caused extreme pain to patient.Eventually, the physician was able to keep the wire access, while pulling out the catheter with the pigtail still formed and placed a non-boston scientific biliary drain over the wire.The device was completely removed from patient's body and the procedure was completed using an alternate device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17638142
Report Number2124215-2023-43523
Device Sequence Number7618315
Product Code FGE
UDI-Device Identifier08714729454953
UDI-Public08714729454953
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K023870
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2024
Device Model Number32337
Device Catalogue Number32337
Device Lot Number0027180543
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2023
Initial Report FDA Received Date08/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
-
-