• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 4.0/15; COVERED CORONARY STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 4.0/15; COVERED CORONARY STENT Back to Search Results
Model Number 434890
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Shock (2072)
Date of Event 08/11/2023
Type of Reportable Event Death
Additional Manufacturer Narrative
The affected pk papyrus stent systems were not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report) and the product release documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the product release documentation confirmed that the devices were manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the occurrence as both no device- and no procedure-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Event or Problem Description
The intention was to treat a lesion in the proximal lcx by rotablation during which a coronary artery perforation type iii occurred.To stop the bleeding balloon dilation was performed.A pk papyrus us 3.0/20 was introduced into the patients body and inflated with 8 atm.However, the perforation was not sealed successfully (reported separately).Therefore, both a drug eluting stent and the complaint pk papyrus us 4.0/15 were implanted, but still the perforation could not be sealed, despite post-dilation with two balloons.Thereafter, the patient underwent an emergent cabg and passed away due to hemorrhagic shock.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 4.0/15
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key17638691
Report Number1028232-2023-04272
Device Sequence Number5590441
Product Code NIV
UDI-Device Identifier07640130447189
UDI-Public(01)07640130447189(17)2409
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/12/2024
Device Model Number434890
Device Catalogue NumberSEE MODEL NO.
Device Lot Number08224910
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age82 YR
Patient SexFemale
-
-