• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Defibrillation/Stimulation Problem (1573)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A stryker service representative performed an initial evaluation of the customer¿s device and was able to verify the reported issue.After replacing therapy pcb assembly, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
 
Event or Problem Description
A customer contacted stryker to report that their device had logged an event code in the memory which is related to a potential issue of the device to to charge for defibrillation energy.In this state, the device may not be able to provide defibrillation therapy, if it were needed.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
dara friedman
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key17639575
Report Number0003015876-2023-01541
Device Sequence Number18076305
Product Code MKJ
UDI-Device Identifier00883873871768
UDI-Public00883873871768
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001950
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2023
Initial Report FDA Received Date08/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-