| Catalog Number |
644400 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Pain (1994); Discomfort (2330)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Since this event resulted in medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the malfunction would be likely to cause or contribute to a serious injury should it recur.As such, this event meets the definition of a reportable event per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that smartlite pro intro kit slp on the led tip get extremely hot and that it has caused postoperative discomfort, pulpitic discomfort and thus endo-treatments as a result after inserting inlays.The customer reportedly, heat generated would cause massive postoperative, pulpitic complaints after insertion of the inlays and he would have several endo treatments as a result.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: we receive yesterday 06th september the complaint smartlite pro here at the equipment service in radolfzell.I have check light power of the cure tip and it is with 976 mw/cm² inside the specification.Also the illuminate tip is with 12 mw/cm² (low) and 24 mw/cm² (high) inside the specification.The temperature at the cure tip backside is after 4x10 sec.On / 5 sec.Off with 28,5°c inside the specification.The battery capacity of both batteries are with 563 mah and 577 mah inside the specification.So don't can find a technical problem on this smartlite pro.Our repair no.(b)(4).
|
| |
|
Search Alerts/Recalls
|