BAXTER HEALTHCARE CORPORATION REPLACEMENT EM2400 DY DISPLAY; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number EXACTA24DYR
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the display cable of an automated compounding device (acd) was frayed causing a line break error.This occurred during an unspecified process step.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection was performed and found the touchscreen nicked, bezel chipped to base metal, back cover nicked, and power cable punctured with no internal wires exposed.Unknown fluids were found on the display.Functional system level testing was performed and failed initially; when powering on the display was unable to log into windows due to customer change of password.After wiping and reimaging the device was slow and the unit took considerably more time to reimage then normal.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The reported damage was confirmed.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPLACEMENT EM2400 DY DISPLAY
Common Device Name
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17641044
Report Number1416980-2023-04326
Device Sequence Number15391572
Product Code NEP
UDI-Device Identifier00085412609041
UDI-Public(01)00085412609041
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberEXACTA24DYR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2023
Supplement Date Received by Manufacturer09/14/2023
Initial Report FDA Received Date08/29/2023
Supplement Report FDA Received Date09/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-