| Catalog Number |
08791686160 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter questioned the results for 3 patients tested for elecsys total psa (total psa) that did not correspond to their clinical history on a cobas e 411 immunoassay analyzer.On (b)(6) 2023 patient 1 result was 1.3 ng/ml.On (b)(6) 2023 a new sample was obtained and the result was 229 ng/ml.On (b)(6) 2023 patient 2 result was less than 0.01 ng/ml.On (b)(6) 2023 a new sample was obtained and the result was 4.69 ng/ml.On (b)(6) 2023 patient 3 result was 0.02 ng/ml.On (b)(6) 2023 a new sample was obtained and the result was 32.2 ng/ml.
|
| |
|
Additional Manufacturer Narrative
|
|
The e411 analyzer serial number was (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The medical device problem code was updated.The customer replaced the reagent pack and made a new calibrator.Calibration and qc passed.The field service engineer (fse) visited the customer site and confirmed the issue was resolved after the customer replaced the reagent pack and made a new calibrator.
|
| |
|