| Catalog Number |
528235 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "staples were found jamming during use on the patient." no patient harm or injury.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The customer returned one unit of 528235 visistat 35w 6/box for investigation.Visual examination of the returned sample revealed that its staples were properly aligned.The stapler was returned with 14 staples remaining in the cover block, indicating that the customer fired at least 21 staples.The trigger was not returned engaged.Clear evidence of use in the form of biological material was present on the devices.An attempt to fire staples was made by engaging the trigger of the stapler using hand pressure.The first 3 staples did not form or release properly and all had to be manually removed.The stapler was disassembled , and the sample's forming tool was replaced with a lab inventory forming tool.The stapler was fired 3 times.All staples formed and released properly.To simulate insertion into the skin, the remaining staples were fired into the skin pad.All staples formed and inserted properly.The stapler was disassembled, and it was confirmed to have been assembled correctly.Die casting anomalies were observed on the forming tool.No other defects or anomalies were observed with the device.A device history record review was performed and no relevant findings were identified.The ifu for this product states, "to obtain optimum staple closure, the trigger must be squeezed all the way in.".The complaint was confirmed based upon the sample received.Upon functional inspection, staples did not consistently form correctly.The sample was disassembled and die casting anomalies on the forming tool were discovered.No other defects or anomalies were observed.It could not be conclusively determined what caused the staples to misfire.A non-conformance was opened by the manufacturing site to further investigate the issue of wide visistat staplers failing to function properly.The forming tool component is under investigation as part of the non-conformance.Teleflex will continue to monitor and trend on complaints of this nature.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that "staples were found jamming during use on the patient." no patient harm or injury.The patient status is reported as "fine".
|
| |
|