• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC 25K FSI-SLI-10S INSERT,PKD; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC 25K FSI-SLI-10S INSERT,PKD; SCALER, ULTRASONIC Back to Search Results
Catalog Number 80392
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event or Problem Description
In this event it is reported that 25k fsi-sli-10s insert,pkd allegedly has sporadic water flow and get hot during use.No injury.
 
Additional Manufacturer Narrative
Returned qty.1 insert, 80392, 23045-00092387, ge, bul test method / gp005-wi02.Inductance: gauge id# 5349 due: 11/30/2024 result: 3.17 ¿ meets spec ¿ does not meet spec ¿ n/a.Tip condition: - ¿ meets spec ¿ does not meet spec ¿ n/a.Stack condition ¿ meets spec ¿ does not meet spec ¿ n/a.Tested on: g130 gage id# 9625 due date: 06/30/24 set pressure: 35 psi water flow: output rate: 35 psi - ¿ meets spec ¿ does not meet spec ¿ n/a.Spray pattern: - ¿ meets spec ¿does not meet spec ¿ n/a.Water leak: ¿ hp sealing ring ¿ proximal ring ¿ distal ring ¿ seam leak ¿ n/a.Vibration: - ¿ meets spec ¿ does not meet spec ¿ n/a.Grip condition: ¿ meets spec ¿ does not meet spec ¿ n/a.Other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#6112 due date: 12/31/2023.The temperature was 83.9° f, no fault found.The specification states are not to exceed 118.4°f.(per (b)(4)).Alleged event: ¿ confirmed - ¿ not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
25K FSI-SLI-10S INSERT,PKD
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key17641661
Report Number2424472-2023-00402
Device Sequence Number5680218
Product Code ELC
UDI-Device IdentifierD003803921
UDI-PublicD003803921
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033705
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number80392
Device Lot Number00092387
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date08/28/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/28/2023
Supplement Date Received by Manufacturer08/28/2023
Initial Report FDA Received Date08/29/2023
Supplement Report FDA Received Date11/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-