|
WALDEMAR LINK GMBH & CO. KG ENDO-MODEL SL; CONNECTION COMPONENT, INCL. TIBIAL PLATEAU, UNCEMENTED, COCRMO, UHMWPE, ROTATING
|
Back to Search Results |
|
| Model Number |
16-2840/07 |
| Medical Device Problem Codes |
Malposition of Device (2616); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Connection component axis did not properly deploy during surgery.By moving the leg several times (extension, flexion, rotation), the connector reached the desired position in the intracondylar region, but the axis did not deploy.The connector has been exchanged.[translated description from customer].
|
| |
|
Additional Manufacturer Narrative
|
|
The review of the device history record showed no deviations.The product complies with the specifications valid at the time of manufacture.The report is delayed due to original classification of the case as non-reportable.Due to the recent fda inspection we reevaluated the complaint in comparison to similar complaints with different outcomes and therefore determined that it is reportable.We have addressed this observation in capa-23-08.
|
| |
|
Search Alerts/Recalls
|
|
|