Catalog Number 955626
Medical Device Problem Codes Protective Measures Problem (3015); Unexpected Shutdown (4019)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/05/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during continuous renal replacement therapy with a prismax machine, the machine "ended treatment with call service powering off suddenly without restarting itself".The bedside nurse reported treatment was ended without the extracorporeal blood being returned to the patient.It was further reported the patient received a blood transfusion.No additional information is available.
 
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
H10: a service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information added to h6 and h10.H10: the event history log was further reviewed.During review the log showed the system halted operation on (b)(6) 2023 at 21:33 due to a b1897 ¿ replacement scale load cell mismatch.The replacement bag was just changed.The reported condition was not verified; however, the most probable cause was due to the act of closing the scale.This could not be verified without a device analysis.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAX MACHINES
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - BROOKLYN PARK
7601 northland drive
n suite 170
minneapolis MN 55428
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17641844
Report Number3003504604-2023-00022
Device Sequence Number12160247
Product Code KDI
UDI-Device Identifier00085412639499
UDI-Public(01)00085412639499
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K190910
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number955626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2023
Supplement Date Received by Manufacturer10/09/2023
11/08/2023
Initial Report FDA Received Date08/29/2023
Supplement Report FDA Received Date10/31/2023
12/06/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-