| Model Number |
A22201C |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report was not returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that during an endometrial ablation using electrode resectoscope cutting loop 24fr 12 degree, the loop resection end broke off of the electrode inside the patient's uterus.The surgeon was able to retrieve the broken pieces and the electrode immediately with a pair of forceps.It was confirmed that all pieces were removed and confirmed all device pieces removed, through the scope already in use.The procedure was completed with a similar device.There was no patient harm reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was prolonged less than 10 minutes to open another device.The patient's anesthesia or sedation was not extended due to the device malfunction.The reported malfunction did not affect the outcome of the procedure or the condition of the patient.No medical intervention (i.E.Treatment outside the scope of the procedure) was required because of the reported problem.The device was inspected prior to use.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additionally, to provide an update with information received (b5), to provide an update to fields (d9, h3, and h4), and to provide a correction to the initial (g2).The device was evaluated by olympus, and it was confirmed that the cutting loop wire was broken off.Additionally, there were non-reportable (non-pae) defects noted.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the cause for the described issue is excessive force while using the affected device or wrong handling of the affected device.The loop on the distal end of the device can wear out during use and may break, burn or melt.However, the specific root of the reported event could not be identified.Olympus will continue to monitor field performance for this device.
|
| |
|