| Model Number |
M00558420 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf code a041001 captures the reportable event of balloon pinhole in the esophagus.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the esophagus during an esophageal stent implantation procedure performed on (b)(6) 2023.During the procedure, the balloon was leaking liquid and the customer reported the balloon had a hole.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf code a041001 captures the reportable event of balloon pinhole in the esophagus.Block h10: investigation results the returned cre wireguided dilatation balloon was analyzed, and a visual examination found that the balloon and catheter of the device had no damages.Functional analysis was performed, and the balloon was inflated without a problem.However, the balloon would not hold pressure due to a pinhole in the balloon (distal section) located approximately 75 mm from the tip.Microscopic inspection found a pinhole located approximately at 75 mm from the tip of the device.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon pinhole can be confirmed.The conclusion is acceptable because, after the product analysis, it was found that the balloon had a pinhole located approximately at 75 mm from the tip of the device.The pinhole found is likely to have occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during/previous the procedure could have caused the problem found on the distal section of the balloon.Therefore, the most probable root cause is an adverse event related to procedure.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the esophagus during an esophageal stent implantation procedure performed on (b)(6) 2023.During the procedure, the balloon was leaking liquid and the customer reported the balloon had a hole.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|