EXACTECH, INC. EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 13MM; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
|
Back to Search Results |
|
| Model Number |
EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 13MM |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical product: serial number item number and full description, (b)(6).314-13-03 - equinoxe cage glenoid medium, alpha, (b)(6).300-10-15 - equinoxe replicator plate 1.5mm o/s, (b)(6).310-01-44 - equinoxe, humeral head short, 44mm (alpha).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported via a equinoxe shoulder study, that a 63 yo female patient, initial right shoulder implanted on (b)(6) 2019, underwent a revision procedure on (b)(6) 2023, approximately 3 years 11 months post the initial procedure due to aseptic humeral loosening.Date of onset,(b)(6) 2023.The standard humeral stem, replicator plate, torque screw, humeral head, and glenoid were removed.The subscapularis was repaired, and a biceps tenodesis was performed.The study states the event was definitely related to the devices and the procedure.The outcome indicates continuing.No device returns anticipated due to clinical study guidelines.No further information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: b3, d1/d2a/d2b, h6.Mdr section codes updated/corrected: b, c, d, g.The revision reported may be the result of the humeral loosening as reported.However, the loosening cannot be confirmed and any potential contributions from manufacturing, user, or patient-related contributing factors to the event cannot be evaluated as the devices were not returned for evaluation and no images or radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|