• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS BIOFLO PICC WITH ENDEXO AND PASV VALVE TECHNOLOGY; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS BIOFLO PICC WITH ENDEXO AND PASV VALVE TECHNOLOGY; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number 75019
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
A territory manager reported an end user experienced an issue with a picc from a bioflo 5f dual lumen pasv max barrier kit.The picc hub cracked during tpn infusion.The catheter was removed.The patient did not experience any adverse effects or harm due to this incident.
 
Additional Manufacturer Narrative
The reported device has yet to be returned to the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
Returned for evaluation was one (1) bioflo picc.As received, the female luer lock component of the white/purple valve was confirmed to be cracked.There were no other visual defects noted to the valve.Reference the attached picture of the cracked hub.Visual inspection of the type and location of the crack in the hub luer is indicative of the hub luer having an over-tightened connection, likely in the presence of tpn or other lipid based infusate the customer's complaint description is confirmed for hub cracked.No component molding non-conformances or other damage was observed during visual inspection of the returned sample.The most likely root cause for the cracked female valve housing is due to an over tightened connection with a mating male luer (likely a needleless connector).A contributing factor to the over-tightened connection may be the mechanics/hand placement during infusion access.It is preferable to grasp the needleless connector (nc) when connecting a syringe/tubing to it rather than grasping the pasv valve luer hub while making a connection with the nc.Grasping the pasv valve hub while connecting a syringe/tubing set to the nc can transmit additional torque to the pasv female luer hub.Inadequate flushing of the device lumen may also be a contributing factor.The dhr for the packaging/assembly lots were reviewed for any deviations related to the reported hub crack failure mode.The review confirms that the lots met all material, assembly and performance specifications with no internal ncrs written.Labeling review: the dfu that is supplied in bioflo kits item number contains the following precaution: it is recommended that only luer lock accessories be used with the · bioflo¿ picc with endexo¿ and pasv¿ valve technology.Repeated over-tightening may reduce hub connector life.Do not use hemostats to secure or remove devices with luer lock hub connections.Flushing - recommended procedure 1.Flush the catheter after every use, or at least every seven days when not in use, to maintain patency.Use a 10 ml syringe or larger.2.Flush the catheter with a minimum of 10 ml of sterile normal saline, using a "pulse" or "stop/start" technique.Warning: if using bacteriostatic saline, do not exceed 30 ml in a 24-hour period.3.Disconnect the syringe and attach a sterile end cap to each luer lock hub.Note: this is the recommended flush procedure for this catheter.If using a different procedure than listed above, the use of heparin may be necessary.Follow institutional protocol for catheter flushing.Precaution: incompatible drug delivery within the same lumen may cause precipitation.Ensure that the catheter lumen is flushed following each infusion.Precaution: if resistance is met when flushing, it is recommended that no further attempts be made.Further flushing may result in catheter rupture.Refer to institutional protocol for clearing occluded catheters.Precaution: place a cap on the hub after use to reduce the risk of contamination.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO PICC WITH ENDEXO AND PASV VALVE TECHNOLOGY
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key17644581
Report Number1317056-2023-00117
Device Sequence Number15448356
Product Code LJS
UDI-Device IdentifierH965750191
UDI-PublicH965750191
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K121089
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number75019
Device Lot Number5773497
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2023
Supplement Date Received by Manufacturer09/14/2023
Initial Report FDA Received Date08/29/2023
Supplement Report FDA Received Date09/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-