Catalog Number D9833
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a bentall procedure on (b)(6) 2023 and suture was used.Whe suture thickness written on the surface in the package (5-0) was different from the description inside the package (6-0).No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6.Component code: g07002 - device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: did the sutures were used on patient? please clarify.No.Could you please clarify if the patient suffered from any signs or consequences due to the issue? please provide more information.No.Is it possible to clarify what was the lot defective product slmeuz or slmpjl? slmeuz is the complaint lot number.On the other hand, sample slmpjl was returned but the difference of size label between the surface of the package and inside one was not found.* could you please clarify if this could be a label incorrect as to prevent proper identification of the product or labeling that provides conflicting product information? this could be a label incorrect as to prevent proper identification of the product.*could you please clarify if this could be a product mixed suture issue? no.The product corresponds to the surface label of the package.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Event related to mw # 2210968-2023-06275, mw # 2210968-2023-06277.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 9/22/2023.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.The returned sample revealed that it was received, twelve unopened packets, and four open samples that pertain to product code d9833, lot slmeuz.During the visual inspection of the samples, a mixed product was observed with the primary folder packet (6-0 instead 5-0), and due to this mismatch, a characterization was performed.In conclusion, the suture belongs to diameter prolene 5-0¿, with a suture length of 37¿.The needle is laser drill type, in size 14 mils, silver color, sales type rb-2, ½ circle, (taper pont).These characteristics were consistent with the product code d9833.Based on the information currently available, the wrong and/or mixed product was identified during the investigation of the samples received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-06275, 2210968-2023-06276, & 2210968-2023-06277.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE SUTURE 36"(90CM) 5-0 BLU
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
ciudad juarez 32604
MX   32604
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key17644588
Report Number2210968-2023-06276
Device Sequence Number18137348
Product Code GAW
UDI-Device Identifier10705031035126
UDI-Public10705031035126
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD9833
Device Lot NumberSLMEUZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2023
Supplement Date Received by Manufacturer09/01/2023
Initial Report FDA Received Date08/29/2023
Supplement Report FDA Received Date09/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-