Section a2, a4 and a5: unknown/ asked but not available.Section b3: date of event: unknown, not provided, but the best estimate date is between (b)(6) 2021 and (b)(6) 2023.Section d6a: if implanted, give date: not applicable, unknown/ asked but not available.Section d6b: if explanted, give date: not applicable, unknown/ asked but not available.Section h3: other 81: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Additional information: section d-9: device available for evaluation: yes.Section d-9: device date returned to manufacturer: 29-aug-2023.Section h-3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received with one haptic bent, the other haptic had a portion separate.The lens was cleaned and, no further issues were observed.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Conclusion: the complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|