EXACTECH, INC. NV GXL LINER NEUTRAL, 40MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 130-40-53
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Synovitis (2094); Osteolysis (2377); Joint Laxity (4526)
Date of Event 01/30/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.
 
Event or Problem Description
As reported via legal documentation, a patient had right hip replacement surgery on (b)(6) 2018.They underwent right hip revision surgery on (b)(6) 2023, approximately 4 years 8 months after their initial implantation.The patient claims to have suffered from extreme pain, limping, instability and a revision surgery as a result of this device.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.The patient has filed a short-form complaint in a multidistrict litigation titled in re: exactech polyethylene orthopedic products liability litigation, mdl no.3044, and pending in the eastern district of new york.Per the transfer order creating this multidistrict litigation, ¿plaintiffs¿allege that their knee or hip replacement devices¿failed prematurely because of degradation of the device¿s polyethylene component, which resulted in the premature removal (or planned removal) of the prosthesis at issue.¿ because the patient has filed a suit in this multidistrict litigation, the patient appears to allege that the patient was injured as a result of premature wear of an exactech polyethylene device.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.D10 concomitants: 170-40-00 - biolox delta femoral 40 mm od, +0mm 5314873; 180-65-25 - alteon 6.5 mm screw, 25 mm 5403711; 180-65-30 - alteon 6.5 mm screw, 30 mm 4763932; 186-01-58 - integrip cc, cluster 58 mm, g3 5396926; 188-00-11 - wedge plasma s/o sz 11 4597715.Based on the available information, the patient involved in incident meets the following risk criteria for early prosthesis wear: combination of largest available femoral head and/or thinnest available liner was used, implanted with a component having a shelf age of greater than 2 years.The most likely underlying cause for the revision reported is a combination of risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINER NEUTRAL, 40MM ID, GROUP 3 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523782617
MDR Report Key17646145
Report Number1038671-2023-02087
Device Sequence Number7979143
Product Code JDI
UDI-Device Identifier10885862207098
UDI-Public10885862207098
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121392
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2020
Device Catalogue Number130-40-53
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/18/2023
Supplement Date Received by Manufacturer01/02/2025
Initial Report FDA Received Date08/29/2023
Supplement Report FDA Received Date01/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
Patient SexMale
-
-