• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA NEONATAL AND PEDIATRIC TTS TRACHEOSTOMY TUB; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA NEONATAL AND PEDIATRIC TTS TRACHEOSTOMY TUB; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Catalog Number 67SN030
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other, other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that during pre-testing the balloon would not inflat.No adverse patient effects were reported by the customer.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer and h6.Evaluation codes: updated.Device evaluation: one sample was returned.During the visual inspection, no defects were found.Based on the analysis performed on the sample provided, no root cause could be determined since the complaint was not confirmed as the samples provided do not show the failure mode reported.The units work as expected and no failure was identified.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.This issue will continue to be monitored and further actions taken accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA NEONATAL AND PEDIATRIC TTS TRACHEOSTOMY TUB
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key17646153
Report Number3012307300-2023-08470
Device Sequence Number7816207
Product Code JOH
UDI-Device Identifier15021312000245
UDI-Public15021312000245
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K914088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2024
Device Catalogue Number67SN030
Device Lot Number3897881
Was Device Available for Evaluation? No
Date Returned to Manufacturer08/22/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2023
Supplement Date Received by Manufacturer09/30/2023
Initial Report FDA Received Date08/29/2023
Supplement Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/27/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-