MDT SOFAMOR DANEK PUERTO RICO MFG INFINITY OCCIPITOCERVICAL UPPER THORACIC SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
| Model Number |
G360T0410 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
07/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider via a manufacturer representative regarding a patient having o-th3 fixation for metastatic spinal tumor.It was reported that there was backout of the oc-plate.Patient had pain during movement.There was no delay in the overall procedure.There was no other patient symptom reported.There were no further complications reported regarding the event.The screws were also backed out.The plate and screws were explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: radiographic image review results: ap x-ray o-t fusion, badly rod construct, lateral x-ray, occipital plate and screws not fastened to skull, and ct reconstruction of ap view of construct is difficult to see hardware failure in the ap view.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|