| Model Number |
AG601261A |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Scar Tissue (2060); Skin Infection (4544)
|
| Date of Event |
08/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Following internal capa opened at inmode, voluntary reports previously submitted by inmode to fda were reviewed retrospectively.Previous voluntary report (mw5092685) was re-assessed and is reported now again as mandatory.Technical inspection of the device and the handpiece revealed no technical issues.Investigation attributed the root cause to a user error: incorrect technique, not congruent with ifu, causing the burns.Contamination was due to lack of aseptic conditions post treatment and is not related to the device.The physician was retrained.
|
| |
|
Event or Problem Description
|
|
Burns on face with subsequent contamination.Residual indentations 3 months post treatment.
|
| |
|