Model Number MAF-TM2
Medical Device Problem Code Device Reprocessing Problem (1091)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the airway mobilescope was reprocessed with an oer-6 endoscope reprocessor that could not drain water sufficiently when replacing the water filter.The water filter was not changed every month and it was operated for about half a year.The issue was found during reprocessing.There were no reports of patient harm.This report is related to the following linked patient identifiers: (b)(6).
 
Additional Manufacturer Narrative
This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the incorrect reprocessing was use of a different procedure than the instruction manual.The event can be detected/prevented by following the instructions for use which state: ¿ ¿chapter 7 section 3 replacing the water filter (maj-2317)¿, the water filter failed to have been correctly replaced.The replacement frequency in the subject facility is approximately every half a year by contrast to every 2 months of replacement frequency which was recommended in the instruction manual.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRWAY MOBILESCOPE
Common Device Name
AIRWAY MOBILESCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17649036
Report Number9610595-2023-12452
Device Sequence Number17984807
Product Code EOQ
UDI-Device Identifier04953170407147
UDI-Public04953170407147
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201300
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAF-TM2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2023
Supplement Date Received by Manufacturer09/14/2023
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date09/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/15/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-