• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL HEDSTROEM M-ACCESS 25MM 15-40; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL HEDSTROEM M-ACCESS 25MM 15-40; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A16MA02590012
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 05/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
In this event it is reported that a hedstroem m-access 25mm 15-40 file broke during use.The broken part was not retrieved.No injury.
 
Additional Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
 
Additional Manufacturer Narrative
Summary: involved product that broke during use was not returned, and cannot be identified and analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review.Root causes are not identified.We will track this kind of event and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEDSTROEM M-ACCESS 25MM 15-40
Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key17650058
Report Number8031010-2023-00835
Device Sequence Number7278518
Product Code EKS
UDI-Device IdentifierJ003A16MA025900121
UDI-PublicJ003A16MA025900121
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA16MA02590012
Device Lot Number1491521
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date08/17/2023
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/17/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date09/05/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-