• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRGAS THERAPEUTICS LLC ISPAN GAS TANK REGULATOR; REGULATOR, PRESSURE, GAS CYLINDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRGAS THERAPEUTICS LLC ISPAN GAS TANK REGULATOR; REGULATOR, PRESSURE, GAS CYLINDER Back to Search Results
Catalog Number 8065797303
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that ophthalmic gas tank valve not measuring the volume contained and not releasing the gas for use, even with a full torpedo.The procedure details and patient impact were not reported.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received scheduled procedure was cataract and vitrectomy combination.
 
Additional Manufacturer Narrative
A check of the batch production record for this lot showed no unusual manufacturing issues.A check of the complaint records showed one other complaint against this lot.A check of confirmed complaints for regulators with low or no flow showed 25 complaints since the beginning of 2016.The sample was not returned.Testing could not be performed.There was no sample returned for this investigation.With no additional, related information provided, the customer reported event was not confirmed.Based upon the information obtained, the root cause of the reported event cannot be determined.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISPAN GAS TANK REGULATOR
Common Device Name
REGULATOR, PRESSURE, GAS CYLINDER
Manufacturer (Section D)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer (Section G)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17650876
Report Number2518435-2023-00029
Device Sequence Number6285710
Product Code CAN
UDI-Device Identifier00380657973033
UDI-Public00380657973033
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2026
Device Catalogue Number8065797303
Device Lot Number105514
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/16/2023
Supplement Date Received by Manufacturer08/31/2023
11/14/2023
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date09/28/2023
12/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-