• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIPATH IMAGING, INC BD TOTALYS SLIDEPREP; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIPATH IMAGING, INC BD TOTALYS SLIDEPREP; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED Back to Search Results
Catalog Number 491346
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using bd totalys slideprep, there was cross-contamination.No patient impact reported.The following information was provided by the initial reporter: "the wad touched the glass, and some pieces were suspected of being contaminated.Was the contamination noticed prior to any results being generated? >> yes.No patient impact.".
 
Additional Manufacturer Narrative
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H.6 investigation summary: complaint reports contamination on slideprep (catalog number 491346) serial number (b)(6).Complaint alleges quad bundles touched the slides and possible contamination.Service adjusted the height of the quad arm, and no contamination reported.Post intervention the instrument was left operating normally.Root cause attributed to quad arm height.This complaint is a confirmed failure of the instrument based on the service investigation.Review of device history record for instrument serial number, (b)(6) is not required because this complaint does not allege an early life failure or failure at installation and has changed configuration since release from manufacturing due to service repairs/pms.Device was installed on (b)(6)2016.Service history review was performed for the instrument (b)(6), and no additional work orders were observed for the complaint failure mode reported.Review of risk management files confirms there are no new or modified risks associated with this failure mode.There is no corrective and preventative action plan in place.
 
Event or Problem Description
It was reported that while using bd totalys slideprep, there was cross-contamination.No patient impact reported.The following information was provided by the initial reporter: "the wad touched the glass, and some pieces were suspected of being contaminated.Was the contamination noticed prior to any results being generated? >> yes.No patient impact.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TOTALYS SLIDEPREP
Common Device Name
PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED
Manufacturer (Section D)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer (Section G)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17651680
Report Number3008007472-2023-00025
Device Sequence Number6257176
Product Code MKQ
UDI-Device Identifier00382904913465
UDI-Public00382904913465
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P970018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number491346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2023
Supplement Date Received by Manufacturer10/27/2023
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date11/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-