| Catalog Number |
366703 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that during use with bd vacutainer® no additive (z) tubes foreign matter was discovered in tube.This occurred with 6-8 tubes.The following information was provided by the initial reporter: customer is reporting foreign matter in tubes.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6 investigation summary: bd had not received samples, but 1 photo was provided for investigation.The photo was reviewed and the indicated failure mode for foreign matter was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode foreign matter.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during use with bd vacutainer® no additive (z) tubes foreign matter was discovered in tube.This occurred with 6-8 tubes.The following information was provided by the initial reporter: customer is reporting foreign matter in tubes.
|
| |
|
Search Alerts/Recalls
|