SYNTHES GMBH DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC Back to Search Results
Catalog Number 05.001.083
Medical Device Problem Codes Break (1069); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.(b)(4).
 
Event or Problem Description
It was reported from colombia that halfway through an unspecified surgical procedure it was observed that the double hose device was making a sound as if it had an exhaust when the engine exploded.It was not reported if there was a delay in the procedure due to the event.It was reported that an unspecified spare device was available for use and the procedure was successfully completed.It was reported that the hose was left marked with red tape and a case report was filed.There was patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: b5: during subsequent follow-up with the customer, additional information was obtained.The reporter stated that the motor did not present a fault, the fault is with the hose.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.A review of the device history was performed and no non-conformances were detected related to the reported condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key17651714
Report Number8030965-2023-10957
Device Sequence Number20415841
Product Code HBB
UDI-Device Identifier07611819831520
UDI-Public07611819831520
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K093361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2023
Supplement Date Received by Manufacturer08/31/2023
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date09/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/09/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-