Catalog Number 401.882S
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: i was reported that on (b)(6) 2023, the patient underwent an orif surgery with the distal ulna plate for a fracture of the distal end of the right ulna.The screw idled when it was inserted into the 4th combination hole from the most distal hole, therefore the screw was removed.The removed screw was then inserted into the second hole from the distal hole, but it idled again.Therefore, the screw was removed.Another screw with the same size was newly opened and inserted into the same 2nd hole from the distal hole.However, the new screw also idled.At the surgeon¿s discretion, the newly opened screw was not removed and left as it was.The surgery was completed successfully within 30 minutes¿ delay.After surgery, the surgeon commented that he may have inserted the screws in a different direction than he had drilled.The patient status was reported to be stable.This report involves one 2.0mm ti locking scr slf-tpng with stardrive recess 12mm.This is report 1 of 2 for (b)(4).
 
Event or Problem Description
The surgeon commented that there was no problem with bone fixation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM TI LOCKING SCR SLF-TPNG WITH STARDRIVE RECESS 12MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17653216
Report Number8030965-2023-10967
Device Sequence Number15016870
Product Code HRS
UDI-Device Identifier10886982205285
UDI-Public(01)10886982205285
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063049
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number401.882S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/16/2023
Supplement Date Received by Manufacturer09/04/2023
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date09/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - PLATES: TRAUMA.; UNK - SCREWS: TRAUMA.
Patient SexUnknown
-
-