Model Number 105581US
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pump prep after the outflow graft was attached, it was noticed that there was leaking at the point of connection where the dacron met the metal.The outflow graft was removed and replaced.The leaking occurred again.Another outflow graft was opened and attached which leaked as well.A third outflow graft was opened and this time it did not leak.Related mfr # 2916596-2023-06113.
 
Additional Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Additional Manufacturer Narrative
Manufacturer¿s investigation conclusion: evaluation of the returned outflow graft confirmed the reported leaking.The outflow graft was returned in like-new condition.Visual inspection of the outflow graft revealed the surface of the knitted graft appeared shiny, indicating that gelatin was present in the graft.No holes or damage were observed through visual or microscopic inspection.The outflow graft was plugged, and the lumen was filled with water.The graft held water until the water pressure was directed toward the hardware connection then water began to drip from the spot where the graft material meets the metal, reproducing the leak.The remainder of the graft did not leak.The graft was sent to pleasanton for additional testing.A manufacturing analysis task was initiated to address the out of box leaking of the outflow graft during pump assembly.Root cause analysis found that the most probable root cause of the leaking graft was related to the stitching method at the supplier.An external corrective action was issued to the supplier.The relevant sections of the device history records for the outflow graft, lot # 8755250, were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate 3 lvas instructions for use (ifu) rev.C is currently available.Section 5 of this ifu contains instructions for unpacking and preparing the sealed outflow graft.Section 5 also states that during the implant process, a complete backup system (implant kit and external components) must be available on-site and in close proximity for use in an emergency.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 SEALED OUTFLOW GRAFT WITH BEND RELIEF
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key17654858
Report Number2916596-2023-06114
Device Sequence Number17987110
Product Code DSQ
UDI-Device Identifier00813024013266
UDI-Public00813024013266
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/07/2025
Device Model Number105581US
Device Lot Number8755250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2023
Supplement Date Received by Manufacturer05/09/2024
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-