(b)(4).D10 - concomitant devices - zimmer segmental distal femoral xt component size b left, catalog #: 00585004201, lot #: 64178798.Zimmer segmental stem collar 25mm, catalog #: 00585204025, lot #: 65530735.Zimmer segmental fluted straight stem extension 9mm diameter x 130mm length, catalog #: 00585205009, lot #: 65083867.Zimmer segmental polyethylene insert xt size b, catalog #: 00585001296, lot #: 65885958.Zimmer segmental articular surface with segmental hinge post size b 12mm, catalog #: 00585002012, lot #: 64834863.Zimmer segmental precoat non-modular cemented tibial component size 3, catalog #: 00588000302, lot #: 65593919.Zimmer segmental trabecular metal medium tibial cone 31mm x 31mm, catalog #: 00545001331, lot #: 64882572.G2- report source - foreign: event occurred in australia.The complainant has indicated that the product will not be returned to zimmer biomet for investigation as the device was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this patient; please see all reports associated with this event: 0001822565-2023-02344, 0001822565-2023-02346, and 0001822565-2023-02347.H3 other text : investigation incomplete.
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h10.H6 - component code - proposed code is mechanical (04) - stem visual evaluation of pictures provided identified that the femoral component had been removed with part of the femur bone still attached, confirming the reported fracture.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|