BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558420
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block e1: the initial reporter facility name is: (b)(6) hospital.Block h6: imdrf device code a0504 captures the reportable event of balloon leak in the esophagus.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the esophagus during a transgastroscopic balloon dilatation procedure performed on (b)(6) 2023.During the procedure, the physician injected liquid to inflate and found that the pressure gauge could not go up.Then, the physician took out the balloon and injected liquid outside of the patient's body and found that the balloon was leaking liquid.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block e1: the initial reporter facility name is: (b)(6) hospital.Block h6: imdrf device code a0504 captures the reportable event of balloon leak in the esophagus.Block h10: investigation results the returned cre wireguided dilatation balloon was analyzed, and a visual examination found that the balloon and catheter of the device had no damages.Functional analysis was performed, and the balloon was inflated without a problem.However, the balloon would not hold pressure due to a pinhole in the balloon (distal section), located approximately 12 mm from the tip.Microscopic inspection found a pinhole located approximately at 12 mm from the tip of the device.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of a balloon leak can be confirmed.The conclusion is acceptable because, after the product analysis, it was found that the balloon had a pinhole located approximately at 12 mm from the tip of the device.The pinhole found is likely to have occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during/previous the procedure, could have caused the problem found on the distal section of the balloon.Therefore, the most probable root cause is an adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the esophagus during a transgastroscopic balloon dilatation procedure performed on (b)(6) 2023.During the procedure, the physician injected liquid to inflate and found that the pressure gauge could not go up.Then, the physician took out the balloon and injected liquid outside of the patient's body and found that the balloon was leaking liquid.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17656896
Report Number3005099803-2023-04637
Device Sequence Number7123816
Product Code FDF
UDI-Device Identifier08714729339359
UDI-Public08714729339359
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2025
Device Model NumberM00558420
Device Catalogue Number5842
Device Lot Number0029114762
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date08/30/2023
Supplement Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
Patient Weight58 KG
-
-