| Catalog Number |
133S-FX-13-T |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Foreign Body Reaction (1868)
|
| Date of Event |
08/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: inferior tab torn away from tissue expander and exposure.
|
| |
|
Event or Problem Description
|
|
Allergan representative reported "inferior tab that was dislodged in the pocket of the patient and caused it to be exposed with a scab.Upon removal of the tissue expander it was identified that the inferior tab had torn away from tissue expander." tissue expander has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Allergan representative reported "inferior tab that was dislodged in the pocket of the patient and caused it to be exposed with a scab.Upon removal of the tissue expander it was identified that the inferior tab had torn away from tissue expander." tissue expander has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ exposure: unable to observe since it is not related to the device.¿ other-technical: observed 6 o¿clock position suture tab was separated from the device.Additional observations: ¿ the piece of suture tab is missing assessed as inconclusive.¿ total delamination of suture tab at the 6 o¿clock position assessed as adhesive failure.¿ observed an opening on radius assessed as surgical damage.¿ brown particles observed outside the device.No further actions are required as the device was implanted.
|
| |
|