• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY C CALCIUM REAGENT KIT; AZO DYE, CALCIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY C CALCIUM REAGENT KIT; AZO DYE, CALCIUM Back to Search Results
Catalog Number 07P57-20
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A final report will be submitted when the evaluation is complete.Section a1 - patient identifier complete entry = (b)(6).All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer observed falsely elevated calcium results generated on the alinity c processing module for patient samples.The following data was provided (customer¿s reference range 2.25 - 2.75 mmol/l): sample id (b)(4) initial result ran on ac02041 = 3.60 mmol/l, repeat = 3.34 mmol/l, sent to another lab and results =1.96 mmol/l, 2.00 mmol/l aliquot after centrifugation run on different platform at another lab = 1.79 mmol/l, 1.79 mmol/l aliquot after centrifugation run on ac02041 = 1.62 mmol/l, 1.63 mmol/l no impact to patient management was reported.
 
Event or Problem Description
The customer observed falsely elevated calcium results generated on the alinity c processing module for patient samples.The following data was provided (customer¿s reference range 2.25 - 2.75 mmol/l): sample id (b)(6) initial result ran on ac02041 = 3.60 mmol/l, repeat = 3.34 mmol/l, sent to another lab and results =1.96 mmol/l, 2.00 mmol/l aliquot after centrifugation run on different platform at another lab = 1.79 mmol/l, 1.79 mmol/l aliquot after centrifugation run on ac02041 = 1.62 mmol/l, 1.63 mmol/l no impact to patient management was reported.
 
Additional Manufacturer Narrative
Update section d4 - expiration date updated from blank to 2/29/2024, g1.And h4 - device mfg date updated from blank to 3/6/2023.The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available, the customer performed troubleshooting by reanalyzing the samples and aliquots, which generated lower results.Quality control was in range at the time of discrepant patient results.Data and information provided by the customer were reviewed and support the complaint issue.Ticket search by lot indicates that the complaint lot performs as expected.Ticket trending review did not identify any trends.Device history review did not identify any non-conformances or deviations with the complaint lot.Labeling was reviewed and adequately addresses the current issue.Based on the investigation, no systemic issue or deficiency of the alinity c calcium reagent, lot 82153un22, was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C CALCIUM REAGENT KIT
Common Device Name
AZO DYE, CALCIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key17660392
Report Number3002809144-2023-00376
Device Sequence Number17890945
Product Code CJY
UDI-Device Identifier00380740132613
UDI-Public00380740132613
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K062855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Catalogue Number07P57-20
Device Lot Number82153UN22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2023
Supplement Date Received by Manufacturer10/06/2023
Initial Report FDA Received Date08/31/2023
Supplement Report FDA Received Date10/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/06/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, AC02041; ALNTY C PROCESSING MODU, 03R67-01, AC02041
Patient SexUnknown
-
-