• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BALLOON BIPOLAR TEMPORARY PACING ELECTRODE, 5FR 110CM; BALLOON FLOW-ASSISTED BIPOLAR ELECTRODE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BALLOON BIPOLAR TEMPORARY PACING ELECTRODE, 5FR 110CM; BALLOON FLOW-ASSISTED BIPOLAR ELECTRODE CATHETER Back to Search Results
Catalog Number 006172P
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Sepsis (2067); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/09/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that in the online survey a physician stated that the patient experienced perforation (arterial or cardiac), pericardial effusion, tamponade, infection, sepsis prior to the device placement and pericardial effusion, bleeding, bacterial infection in the groin, sepsis complications were directly attributable to the device and fever was needed for infection, and surgery was needed for sepsis, and cabg was needed for pericardial effusion, and anemia led to transfusion was needed for bleeding in relation to temporary pacing electrode.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned, and further investigation was not conclusive.There have been no changes to the materials used or the manufacturing processes involved with this product.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "read this document in its entirety prior to use." "inappropriate electrical connections, e.G.Into a wall socket, may pose serious risk of adverse health consequences or death." "this device should be used only by or under the supervision of physicians trained in the techniques of transvenous intracardiac studies and temporary pacing." "the risks of using temporary pacing catheters include those risks related to heart catheterization, such as thromboembolism, perforation, tamponade, and infection.The induction of an unintended arrhythmia is a known complication." "visually inspect the catheter, under sterile conditions, for kinks in the catheter shaft, integrity of the connector, condition of electrodes, and any other damage." "insertion instructions using a needle cannula: open the package and place the contents on a sterile field.Prep the skin at the site of insertion and inject a local anesthetic.Remove the protective guard from the needle cannula.Enter the vein with the needle cannula.Simultaneous aspiration into a syringe will help confirm vessel entry.Remove the syringe and the needle.If using an open-lumen catheter, flush the catheter with a heparinized solution.Remove any stylette prior to insertion.Using the aid of fluoroscopy or an ecg monitor, advance the catheter through the cannula to the desired position.If using a balloon catheter, inflate the balloon when the catheter is in the right atrium.Please note that the balloon can be inflated or deflated only when the stopcock is parallel to the catheter shaft.Do not pull the catheter back through the cannula as it may cause damage to the catheter.If using a balloon catheter, deflate the balloon after the catheter has reached the desired location.Test the pacing characteristics for optimal pacing.Pull the cannula back and secure it to the proximal end of the catheter.Secure the electrode catheter in place at the insertion site." "electrical connections for measuring intracardiac ecgs: insert adapter pin into standard 2 mm (0.080¿) pin receptacle of equipment.If equipment contains a locking mechanism such as a collet or thumbscrew, tighten down onto adapter.Leave affixed to equipment.Warning: inappropriate electrical connections, e.G.Into a wall socket, may pose serious risk of adverse health consequences or death.Thread leads of catheter through the adapter eyelet.Connect the negative jack (marked ¿distal¿) to the v-lead of the ecg, and the positive jack (unmarked) to the positive terminal of the external pulse generator.Electrical connections for pacing insert adapter pin into standard 2 mm (0.080¿) pin receptacle of equipment.If equipment contains a locking mechanism such as a collet or thumbscrew, tighten down onto adapter.Leave affixed to equipment.Warning: inappropriate electrical connections, e.G.Into a wall socket, may pose serious risk of adverse health consequences or death.Thread leads of catheter through the adapter eyelet.Connect the negative jack (marked distal) to the negative terminal of the external pulse generator, and the positive jack (unmarked) to the positive terminal of the pulse generator." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
 
Event or Problem Description
It was reported that in the online survey a physician stated that the patient experienced perforation (arterial or cardiac), pericardial effusion, tamponade, infection, sepsis prior to the device placement and pericardial effusion, bleeding, bacterial infection in the groin, sepsis complications were directly attributable to the device and fever was needed for infection, and surgery was needed for sepsis, and cabg was needed for pericardial effusion, and anemia led to transfusion was needed for bleeding in relation to temporary pacing electrode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLOON BIPOLAR TEMPORARY PACING ELECTRODE, 5FR 110CM
Common Device Name
BALLOON FLOW-ASSISTED BIPOLAR ELECTRODE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17660867
Report Number1018233-2023-06454
Device Sequence Number12324014
Product Code LDF
UDI-Device Identifier00801741011191
UDI-Public(01)00801741011191
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K800298
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number006172P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2023
Supplement Date Received by Manufacturer01/03/2024
Initial Report FDA Received Date08/31/2023
Supplement Report FDA Received Date01/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-