| Catalog Number |
333512 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b0(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: detailed inquiry description: fx catheter fractured during removal.Upon removal of fx catheter, the polamide coating fractured at approx 5 cm from distal end leaving the wire exposed.Full removal was accomplished just took more time and attentive care to remove the catheter.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for further evaluation.A thorough investigation could not be performed without a sample to evaluate.Therefore, no root cause is able to be determined at this time.Although the tip of the needle was reported to be broken off, since it was not broken or damaged upon the opening of the kit and it had been used in a procedure, per the manufacturer's investigation this defect is not likely to have happened during the manufacturing process.It is believed to happen during application.User will be referred to the ifu which states, " do not apply excessive force during needle advancement.Do not utilize needle if it bends or tip becomes damaged.If resistance is felt during advancement of the needle, carefully correct the orientation of the needle but never apply excessive force." review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|