MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ARTICULEZE M 28MM 1.5; METAL FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ARTICULEZE M 28MM 1.5; METAL FEMORAL HEAD Back to Search Results
Catalog Number 136511500
Medical Device Problem Codes Naturally Worn (2988); Noise, Audible (3273)
Health Effect - Clinical Codes Fatigue (1849); Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Discomfort (2330); Cramp(s) /Muscle Spasm(s) (4521); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Date of Event 08/24/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).D4-the device catalog number is unknown; therefore, udi is unavailable.E3 initial reporter occupation: lawyer.H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.The adverse symptoms alleged and product code reported is associated with the depuy metal on metal articulation.A complaint database search and/or device manufacturing (dhr) reviews will not be performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Pinnacle litigation complaint received ad 17 aug 2023.On (b)(6) 2015 patient underwent a revision surgery due to severe pain, discomfort, inflammation, squeaking, elevated cobalt and chromium levels, friction and wear between the metal head and metal liner caused large amounts of toxic cobalt chromium metal ions.During the procedure, it was further discovered that she had developed significant metallosis around the hip with accompanying tissue damage.Doi: (b)(6) 2008.Dor: (b)(6) 2015.Left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The adverse symptoms alleged and product code reported is associated with the depuy metal on metal articulation.A complaint database search and/or device manufacturing (dhr) reviews will not be performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.H10 additional narrative:.
 
Event or Problem Description
Medical records were received: after review of the medical records the patient was revised to address adverse local tissue reaction metal on metal bearing, elevated metal ions and hip pain.Operative note reported there was a dishwater type fluid from the joint, surgical delay due to metal liner difficult to disengage.On (b)(6) 2023 mri spine, muscle spasm and left leg weakness.However there was no invasive treatment or revision noted.Doi: (b)(6) 2008; dor: (b)(6) 2015: left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, "[pinnacle litigation complaint received ad 17 aug 2023.On (b)(6) 2015 patient underwent a revision surgery due to severe pain, discomfort, inflammation, squeaking, elevated cobalt and chromium levels, friction and wear between the metal head and metal liner caused large amounts of toxic cobalt chromium metal ions.During the procedure, it was further discovered that she had developed significant metallosis around the hip with accompanying tissue damage." the product was not returned to depuy synthes, however photos were provided for review.See attachment ((b)(4) _x-ray_images-received 31-jan-2024).The x-ray investigation revealed that there were no product problems observed with articuleze m 28mm 1.5 the overall complaint was unconfirmed as the observed condition of the articuleze m 28mm 1.5 would not contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULEZE M 28MM 1.5
Common Device Name
METAL FEMORAL HEAD
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17661089
Report Number1818910-2023-17940
Device Sequence Number12314004
Product Code JDI
UDI-Device Identifier10603295033073
UDI-Public10603295033073
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K883460
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2012
Device Catalogue Number136511500
Device Lot Number2512568
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2023
Supplement Date Received by Manufacturer12/27/2023
01/31/2024
02/07/2024
02/26/2024
05/27/2024
Initial Report FDA Received Date08/31/2023
Supplement Report FDA Received Date01/02/2024
02/08/2024
02/12/2024
03/06/2024
05/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/05/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M 28MM 1.5; CORAIL2 NON COL HO SIZE 10; PINNACLE MTL INS NEUT28IDX48OD; UNK HIP ACETABULAR CUP PINNACLE; UNK HIP ACETABULAR CUP PINNACLE; UNK HIP ACETABULAR LINER METAL PINNACLE; UNK HIP FEMORAL STEM CORAIL; UNK HIP FEMORAL STEM CORAIL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight58 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-