| Model Number |
PHALANGEAL REAMER, 20MM |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, it was reported by a sales representative via sems the following: an ar-8944mr-18 metatarsal reamer is getting dull.An ar-8944mr-20 metatarsal reamer is getting dull.An ar-8944mr-22 metatarsal reamer is getting dull.An ar-8944pr-18 phalangeal reamer is getting dull.An ar-8944pr-20 phalangeal reamer is getting dull.An ar-8944pr-22 phalangeal reamer is getting dull.This was discovered during an unspecified procedure, with no adverse event or patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint was confirmed.One unpackaged ar-8944pr-20, serial/batch number (b)(6), was received for investigation.Upon visual evaluation, it was noted signs of wear and tear.The sharp edges at the cutting tip have nicks and are dull.The most likely reason for the reported failure is the wear and tear damage incurred over repeated usage.Devices with cutting functions or sharp points become dull with continuous use.This condition does not indicate a device defect.This condition indicates normal wear.
|
| |
|