| Catalog Number |
08056722190 |
| Medical Device Problem Codes |
High Test Results (2457); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The c503 analyzer serial number is (b)(6).
|
| |
|
Event or Problem Description
|
|
The initial reporter complained of a discrepant high result for 1 patient urine sample tested for tina-quant albumin gen.2 (albt2) on a cobas pro c 503 analytical unit.The initial result was 40.1 mg/l.The repeat result was less than 0.583 mg/l with a data flag.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration and qc were acceptable.The low albt2 result was believed to be correct.The customer did not provide any additional information for the investigation.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|