ROCHE DIAGNOSTICS ALBT2; ALBUMIN IMMUNOLOGICAL TEST SYSTEM Back to Search Results
Catalog Number 08056722190
Medical Device Problem Codes High Test Results (2457); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The c503 analyzer serial number is (b)(6).
 
Event or Problem Description
The initial reporter complained of a discrepant high result for 1 patient urine sample tested for tina-quant albumin gen.2 (albt2) on a cobas pro c 503 analytical unit.The initial result was 40.1 mg/l.The repeat result was less than 0.583 mg/l with a data flag.
 
Additional Manufacturer Narrative
Calibration and qc were acceptable.The low albt2 result was believed to be correct.The customer did not provide any additional information for the investigation.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALBT2
Common Device Name
ALBUMIN IMMUNOLOGICAL TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17662097
Report Number1823260-2023-02841
Device Sequence Number14992631
Product Code DCF
UDI-Device Identifier07613336120965
UDI-Public07613336120965
Combination Product (Y/N)Y
Initial Reporter CountryTC
PMA/510(K) Number
K101203
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number08056722190
Device Lot Number617156
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/12/2023
Supplement Date Received by Manufacturer09/15/2023
Initial Report FDA Received Date08/31/2023
Supplement Report FDA Received Date09/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-