| Model Number |
QUAD TENDON STRIPPER- CUTTER |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 8/2/2023, it was reported by a sales representative via sems that an ar-2384 quad tendon stripper-cutter, part of set ar-2382s with lot number 15100379, had an issue.During the case, the surgeon noticed that the blades on the quad tendon stripper-cutter were dull.No patient affect.This occurred during use in an unspecified procedure with no patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.Upon visual evaluation, it was noted that the edges of the quad tendon stripper-cutter are worn, showing nicks, dents, and signs of having been used heavily.Dfu-0023-eo warns against using dull devices.2.Devices with cutting functions or sharp points become dull with continuous use.This condition does not indicate a device defect.This condition indicates normal wear.Dull devices may require replacement if they no longer perform as designed.Inspection prior to use should include verifying the cutting ability and sharpness of these points and edges.The most likely reason for the reported failure is the wear and tear damage incurred over repeated usage.
|
| |
|
Search Alerts/Recalls
|