• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. QUAD TENDON STRIPPER- CUTTER; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. QUAD TENDON STRIPPER- CUTTER; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number QUAD TENDON STRIPPER- CUTTER
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 8/2/2023, it was reported by a sales representative via sems that an ar-2384 quad tendon stripper-cutter, part of set ar-2382s with lot number 15100379, had an issue.During the case, the surgeon noticed that the blades on the quad tendon stripper-cutter were dull.No patient affect.This occurred during use in an unspecified procedure with no patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Complaint is confirmed.Upon visual evaluation, it was noted that the edges of the quad tendon stripper-cutter are worn, showing nicks, dents, and signs of having been used heavily.Dfu-0023-eo warns against using dull devices.2.Devices with cutting functions or sharp points become dull with continuous use.This condition does not indicate a device defect.This condition indicates normal wear.Dull devices may require replacement if they no longer perform as designed.Inspection prior to use should include verifying the cutting ability and sharpness of these points and edges.The most likely reason for the reported failure is the wear and tear damage incurred over repeated usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUAD TENDON STRIPPER- CUTTER
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17662127
Report Number1220246-2023-07699
Device Sequence Number6324504
Product Code MDM
UDI-Device Identifier00888867098244
UDI-Public00888867098244
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQUAD TENDON STRIPPER- CUTTER
Device Catalogue NumberAR-2384
Device Lot Number012316
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2023
Supplement Date Received by Manufacturer08/02/2023
Initial Report FDA Received Date08/31/2023
Supplement Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/26/2023
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-