• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP RVS CNTRL 6.5X50MM ST/RST; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP RVS CNTRL 6.5X50MM ST/RST; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES Back to Search Results
Catalog Number 115400
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that while the surgeon was implanting the implant during a shoulder arthroplasty, the implant fractured.The surgeon was not able to remove all of the implant and the patient retained a portion of the implant.The surgeon was able to implant a second implant.Multiple attempts have been made to obtain additional information.No response has been received at this time.
 
Additional Manufacturer Narrative
(b)(4).H3: customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that while the surgeon was implanting the implant during a shoulder arthroplasty, the implant fractured.The surgeon was not able to remove all of the implant and the patient retained a portion of the implant.The surgeon was able to implant a second implant.There are no plans to revise the patient and remove the piece of the implant that was retained in the patient's bone.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were not provided.The root cause of the reported issue is attributed to user-error.Per comprehensive® vault reconstruction system surgical technique and the comprehensive® reverse shoulder system surgical technique, "insert the desired length 6.5 mm central screw and completely tighten with the 3.5 mm hex driver." the central screws are to be inserted using a driver without power.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVS CNTRL 6.5X50MM ST/RST
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17663468
Report Number0001822565-2023-02382
Device Sequence Number17341104
Product Code PHX
UDI-Device Identifier00880304677128
UDI-Public(01)00880304677128(17)320308(17)653590
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K193373
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/09/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number115400
Device Lot Number653590
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2023
Supplement Date Received by Manufacturer10/09/2023
01/05/2024
Initial Report FDA Received Date08/31/2023
Supplement Report FDA Received Date11/01/2023
01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age60 YR
Patient SexMale
-
-