| Model Number |
CORKSCREW FT, BC, SUTURETAPE, 4.75M |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 8/4/2023, it was reported by a facility representative via email that (2) ar-3632sp fibertak sp and an ar-1927bct-475 biocomposite corkscrew broke during insertion.This was discovered during a procedure on (b)(6) 2023.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely causes of the reported failure are improper bone preparation, misaligned insertion, and prying/leveraging the device during insertion.
|
| |
|