Catalog Number 1009-00
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have received the complaint device for evaluation and confirmed the reported incident.When the green handle was pulled to close the blades into its retainer slot one of the blades was protruding out of the retainer slot and didn't sit properly on the slot.No damage to the blade was found.Previous investigations into this issue have found the failure was determined to be centering hoop to wire welding error along with operator adjustment error.During the welding process, this centering hoop was likely not properly aligned against its retainer slot.The blade process includes manual manipulation of each blade by the operator.Operator errors are possible due to the manual nature of the adjustments.We have conducted a lot history review and did not find any issues noted in the manufacturing or packaging process that could be related to this issue.All qc tests passed their requirements.Further, we have not received any other complaints of a similar nature for devices from this lot.Capa 2022-025 was opened to further investigate this issue.
 
Event or Problem Description
It was reported the lemaitre valvulotome did not sheath properly after first pass during insitu bypass- fem-tib procedure.A new valvulotome was opened and used for the 2nd pass.No injury was reported to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEMAITRE VALVULOTOME
Common Device Name
VALVULOTOME
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
7814251683
MDR Report Key17664584
Report Number1220948-2023-00140
Device Sequence Number7087713
Product Code MGZ
UDI-Device Identifier00840663106653
UDI-Public(01)00840663106653
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K140042
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 08/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1009-00
Device Lot NumberELVH2239V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2023
Initial Report FDA Received Date08/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-