MAUDE Adverse Event Report: SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD Back to Search Results
Model Number 7070M-100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 12/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique guide, ifu, certificates of conformance and fmea, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause for the fracture could not be determined.Bone fracture is a known risk of placement of metallic devices into pelvic bone.
 
Event or Problem Description
In (b)(6) 2023, si-bone learned of an ilium fracture case that had occurred in late 2021.The patient had multilevel spine fusion surgery with placement of both pelvic fixation screws and ifuse in the sai configuration on (b)(6) 2021.About 2 weeks post-op, the patient heard a pop.Ct scan shows a large, displaced ilium fracture.The surgeon installed screws a few month later to help fixate the ilium fracture.It is unknown whether the ifuse implants were causally related, but it was the implant that was closest to the fracture location and patient movement can impose a large lever force on the pelvic bones.Bone fracture is a known risk of placement of metallic devices into pelvic bone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
Manufacturer (Section G)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
Manufacturer Contact
w. carlton reckling
471 el camino real
suite 101
santa clara, CA 95050-4482
4082070700
MDR Report Key17666938
Report Number3007700286-2023-00001
Device Sequence Number6325162
Product Code OUR
UDI-Device Identifier00851085007541
UDI-Public00851085007541
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K193524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/14/2026
Device Model Number7070M-100
Device Lot Number9046141
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2023
Initial Report FDA Received Date08/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age64 YR
Patient SexMale
-
-