The device was returned to olympus for evaluation and the customer's allegation was confirmed, the device evaluation found that there was no image.Additional findings include the following: the instrument channel was leaking, the objective lens was missing cement on the lens, the light guide lens was missing cement on the lens, the insertion tube had a heavy dent, and the light guide tube was determined to be a third party (non-olympus) tube.The investigation is ongoing, a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.It has been over 20 years since the subject device was manufactured.Based on the results of the investigation, it is likely while the user was handling the device, physical stress was applied to insertion section which led to damage of ccd-unit (charge-coupled device unit).Also, it is likely while the user was handling the device, biopsy channel leaked, which led to water invasion of the device, then abnormality of electronic parts.As a result, no image event however, a definitive root cause could not be determined.The following information is stated in the instructions for use (ifu): ¿important information ¿ please read before use precaution for disappeared or frozen endoscopic image; warning ·follow the precautions given below when handling the instrument.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.Precaution for disappeared or frozen endoscopic image; caution ¿ turn the video system center off before connecting or disconnecting the videoscope cable from the electrical connector on the endoscope.Turn the switch on or off only when the videoscope cable is connected to both the video system center and electrical connector on the endoscope.Failure to do so can result in equipment damage, including destruction of the ccd.¿ do not touch the electrical contacts inside the electrical connector.Ccd damage may result.3.6 inspection of the endoscopic system inspection of the endoscopic image 4.While observing the palm of your hand, confirm that the wli and nbi endoscopic image is free from noise, blur, fog, or other irregularities.5.Angulate the bending section and confirm that the endoscopic image is free from momentary disappearing or other irregularities.Chapter 5 troubleshooting if the endoscope is visibly damaged, does not function as expected or is found to have irregularities during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope.Contact olympus.Some problems that appear to be malfunctions may be correctable by referring to section 5.1, ¿troubleshooting guide¿.If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to olympus for repair.Olympus does not repair accessory parts.If an accessory part becomes damaged, contact olympus to purchase a replacement.¿ olympus will continue to monitor field performance for this device.
|