| Model Number |
ND2110X15B |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a ventilator was alarming circuit breaker for circuit failure.There was no harm or injury reported.The device has not been returned to the manufacturer.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 other text : not returned to the manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported receiving information alleging a ventilator was alarming circuit breaker for circuit failure.There was no harm or injury reported.The ventilator was evaluated by a third party service center and the customer¿s complaint was confirmed.The device's proportional valve and the 3 way solenoid valve were replaced to address the issue.
|
| |
|