| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It war reported the trial fractured during the proecedure.There was no patient harm or impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information. h6-component code- suggested code: mechanical (g04) - provisional bottom.Visually verified item lot combination and reviewed manufacturing date as tasp exhibits signs of repeated use (nicks, gouges) and the post feature has fractured off, not all pieces returned.The device history records & receiving inspection report were reviewed for deviations and / or anomalies with no deviations / anomalies identified.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Medical records were not provided.Fractured tasps were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.Complaint is confirmed via product return & evaluation.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|